RTD Temperature Calculator: Formula & How to Use

Card No. 06·1238 · Updated · Standard formula, shown below

The calculator on this card is switched off while it is checked. The formula is below, and the worked example shows how to do the sum by hand.

An RTD temperature calculator converts the resistance reading of a platinum resistance thermometer (like a PT100) into a temperature in °C or °F. You enter the measured resistance and the RTD type (e.g., PT100, PT1000), and the calculator returns the corresponding temperature using the standard Callendar-Van Dusen equation. This is essential for anyone working with industrial sensors, lab equipment, or HVAC systems where accurate temperature measurement matters.

Formula

The calculator uses the linear approximation of the Callendar-Van Dusen equation for temperatures from 0°C to 850°C:

T = (RT – R0) / (α · R0)

  • RT – measured resistance at temperature T (Ω)
  • R0 – nominal resistance at 0°C (100 Ω for PT100, 1000 Ω for PT1000)
  • α – temperature coefficient, typically 0.00385 Ω/Ω/°C for standard platinum (DIN EN 60751)

For the PT100, the denominator α·R0 = 0.385 Ω/°C. So the formula simplifies to T = (R – 100) / 0.385. The same logic applies to PT1000 with R0 = 1000 Ω.

Worked Example

Using the default values from the calculator: measured resistance = 138.5 Ω, RTD type = PT100.

  1. Subtract R0 from the measured resistance: 138.5 – 100 = 38.5 Ω
  2. Divide by α·R0 (0.385 Ω/°C): 38.5 / 0.385 = 100°C
  3. Result: 100.00°C

This matches the standard PT100 table where 138.5 Ω corresponds to 100°C.

How RTD Sensors Work

RTDs (Resistance Temperature Detectors) exploit the fact that the electrical resistance of platinum changes predictably with temperature. Unlike thermocouples, which generate a voltage, RTDs require an external current source to measure resistance. The sensing element is usually a thin film or wire-wound platinum resistor. The relationship between resistance and temperature is nearly linear, but for high accuracy (better than ±0.5°C), you need the full Callendar-Van Dusen equation, which adds a quadratic term for negative temperatures and a cubic term for positive ones.

Choosing the Right RTD Type

Two common types are PT100 and PT1000. The number indicates resistance at 0°C. PT100 is the most widespread in industry, while PT1000 is preferred in battery management systems and HVAC because its higher resistance reduces the effect of lead wire resistance (which can be significant over long distances). If you are measuring over long cable runs, a 3-wire or 4-wire configuration is recommended to cancel lead resistance. For most DIY projects, a PT100 with a simple 2-wire hookup is fine if the leads are short.

Common Problems and Misconceptions

One common mistake is using the linear formula below 0°C, where it deviates by several degrees. For example, at -50°C, a PT100 actually reads about 79.7 Ω, but the linear formula gives (79.7-100)/0.385 = -52.7°C – an error of 2.7°C. Always use the full equation for sub-zero temperatures. Another pitfall is ignoring self-heating: the excitation current warms the sensor slightly, causing a high reading. Keep the current below 1 mA for thin-film elements. Also, beware of using the wrong α coefficient – some older sensors use 0.003902 instead of 0.00385, leading to a systematic error of about 1.3%.

Accuracy and Calibration

The intrinsic accuracy of an RTD depends on its class. Class A PT100 has an accuracy of ±0.15°C at 0°C, while Class B is ±0.3°C. However, the measuring circuit (ADC, reference resistor, lead resistance) often introduces larger errors. For critical applications, perform a two-point calibration at known temperatures (e.g., ice point and boiling point) and adjust the offset and gain. The calculator gives you the ideal temperature for a given resistance, but real-world sensors may deviate slightly.

Alternatives to RTD

Thermocouples are cheaper and handle higher temperatures (up to 1700°C) but are less accurate and require cold-junction compensation. Thermistors have higher sensitivity but a very nonlinear response and limited range (typically -50 to 150°C). Integrated digital sensors like the DS18B20 are easy to use but have a narrower range and lower accuracy. Choose an RTD when you need stability, repeatability, and accuracy over a moderate range (-200 to 850°C).

Frequently Asked Questions

How do I convert PT100 resistance to temperature manually?

Use the linear formula T = (R – 100) / 0.385 for temperatures above 0°C. For example, if you measure 119.4 Ω, then T = (119.4 – 100) / 0.385 = 50.4°C. For negative temperatures, use the full polynomial to avoid errors.

What is the difference between PT100 and PT1000?

The nominal resistance at 0°C: 100 Ω vs 1000 Ω. PT1000 has higher sensitivity (3.85 Ω/°C vs 0.385 Ω/°C) and is less affected by lead resistance, making it better for low-power or remote sensing. PT100 is more common in industrial panels and has more interchangeable parts.

Can I use a multimeter to measure RTD resistance?

Yes, but the multimeter’s test current may cause self-heating, and the lead resistance of the meter can add error. For best results, use a 4-wire ohmmeter or a dedicated RTD transmitter. For a quick check, a good multimeter on the ohms range is acceptable for rough readings.

Why does my RTD read higher than the actual temperature?

Possible causes: self-heating (too high excitation current), lead wire resistance (if using 2-wire), or sensor drift due to contamination or mechanical strain. Check the wiring and reduce the current to below 1 mA. Also verify that you are not using a 3-wire sensor in a 2-wire configuration without compensation.

What is the maximum temperature for a PT100?

For standard thin-film PT100, the range is -50°C to +500°C. Wire-wound elements can go up to 850°C, but the sheath material and connection method limit the practical maximum. Always check the sensor’s datasheet for the specific range.

How do I choose between 2-wire, 3-wire, and 4-wire RTD?

2-wire is simplest but includes lead resistance in the measurement. 3-wire cancels lead resistance if the two leads are identical. 4-wire gives the most accurate measurement by using separate sense and source leads. For long cables or high accuracy, use 3- or 4-wire.

Related

Formula used
in r | Measured resistance | Ω | 138.5 | 0 | 1000
sel rtd | RTD type | PT100=100, PT1000=1000
out temp | Temperature | °C | (r - sel) / 0.385 | 2

Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work

Method: the standard formula for this calculation, rounded as each result states. See how the calculators work or report an error.